Command Reference
R&S
®
ZNL/ZNLE
652
User Manual 1178.5966.02 ─ 07
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
11
.
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC:FEED 'CH4TR1'
Display the generated trace in diagram area no. 2.
Usage:
Setting only
Manual operation:
Table 10-7: String identifiers for measurement results
Note:
All port numbers in a result identifier refer to
logical (=DUT) ports
; to avoid ambiguities they must be represented by the same
number of digits (e.g. S21 or S0201). The valid port numbers are determined by the channel's logical port configuration
For details about the measurement results see
Chapter 8.3, "Measurement Results"
'S11' | 'S12' | ... | 'S0101' | ...
Single-ended S-parameters S<i><j> for logical (DUT) ports <i> and <j>. To
avoid ambiguities, <i> and <j> must be specified with the same number of
digits.
'SCD11' | ...
S-parameters involving balanced ports must be specified in the form
S<m_i><m_j><i><j>, where <m_i> and <m_j> denote the port modes of the
related logical ports <i> and <j>. In general, for the port modes <m_i><m_j>
all pairs of D (differential, balanced), C (common, balanced) and S (single-
ended, unbalanced) are allowed.
'Y11' | ... | 'YSS11' | ... | 'YCC11' | ... | 'YDD11' | 'Z11'
| ... | 'ZSS11' | ... | 'ZCC11' | ... | 'ZDD11' | ...
Short-circuit Y-parameters and open-circuit Z-parameters with port modes
and port numbers like for normal mode S-parameters.
Selecting a parameter Y...<n><m> or Z...<n><m> sets the range of logical
port numbers to be considered for the Y and Z-parameter measurement to
<n>:<m>.
'Y-S11' | ... | 'Y-SSS11' | ... | 'Y-SCC11' | ... | 'Y-
SDD11' | 'Z-S11' | ... | 'Z-SSS11' | ... | 'Z-SCC11' | ...
| 'Z-SDD11' | ...
S-parameters converted to matched-circuit admittances and impedances
with port modes and port numbers like for normal mode S-parameters.
'SHUNT-S21'
Converted impedances measured using shunt-thru method. Use
to set the probe tip impedance.
'A1D2' | ... | 'A01D02' | ... | 'B1D2' | ... | 'B01D02' | ...
Wave quantities A<meas>D<drive> (for a wave) and B<meas>D<drive> (for
b wave) for logical ports <meas> and <drive>.
'A1' | ... | 'A01' | ... | 'B1' | ... | 'B01' | ...
Wave quantities A<meas>D1 (for a wave) and B<meas>D1.
'B2/A1' | ... 'B02/A01' | ...
Ratio of wave quantities with port numbers like for normal mode S-parame-
ters.
'B2D1/A1D1' | ... 'B02D01/A01D01' | ...
Ratios of wave quantities.
'IMB1-23' | 'IMB23-1' | 'IMB1-24' | ...
Differential Imbalance parameters between a balanced and two single-ended
logical ports
'KFAC21' | 'KFAC12' | ...
Stability factor K (for unbalanced ports only)
'MUF121' | 'MUF112' | ...
Stability factor 1 (for unbalanced ports only)
'MUF221' | 'MUF212' | ...
Stability factor 2 (for unbalanced ports only)
VNA Command Reference