Active
The checkbox in the "Active" column activates/deactivates the configured length and
loss parameters for the respective
(i.e. adds/removes them to/from the calculation
flow) without changing the parameter values.
Remote command:
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:COMPensation[:STATe]
Reset Offsets
The "Reset Offsets" button resets the length and loss parameters for all ports to their
default values.
Remote command:
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>[:STATe]
All Offsets On / All Offsets Off
Activates/deactivates the length and loss compensation for all ports.
Use the checkboxes in the
"Active" column to activate/deactivate the length and
loss compensation for selected ports.
Remote command:
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:COMPensation[:STATe]
9.14.1.3
Single Ended Panel
Allows you to activate or deactivate single ended deembedding/embedding for
selected physical ports. For the "2-Port Data" network type, it is also possible to
change the underlying
s2p
Touchstone file from the dock widget.
Access:
> "Single Ended"
The "Single Ended" panel can also be activated by selecting the
softtool tab. Refer to its description for background information, parameters and addi-
tional remote commands.
Active
The checkbox in the "Active" column activates or deactivates the selected "Single
Ended" de-/embedding
(i.e. adds or removes it to/from the calcula-
tion flow) without changing its parameters.
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>[:
STATe]
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>[:
STATe]