9.11.1.1
Controls on the Peak Tab
Max / Min
Sets the active marker to the absolute maximum or minimum in the search range, i.e.
to the largest or smallest of all response values. If a complex trace format (e.g. a polar
diagram) is active, the marker is set to the measurement point with the maximum or
minimum magnitude.
"Max" and "Min" also overwrite the current "Search Mode" (--> "Search Min" and
"Search Max") and the "Peak Type" for the peak search functions.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
MINimum | MAXimum
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
Center = Marker
See
"Center = Marker / Start = Marker / Stop = Marker / Span = Marker"
Next Peak
Sets the active marker to the next local maximum or minimum in the search range,
depending on the selected
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
NPEak
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
Peak Left / Peak Right
Sets the active marker to the next local maximum or minimum to the left or right of the
current marker position, depending on the selected