Command Reference
R&S
®
ZNL/ZNLE
580
User Manual 1178.5966.02 ─ 07
Event
Switching on
supply voltage
Power-On-Status-
Clear
DCL, SDC
(Device
Clear,
Selected
Device Clear)
*RST or
SYS-
Tem:PRE-
Set:ALL
STA-
Tus:PRE-
Set
*CLS
Effect
0
1
Clear STB,ESR
yes
yes
Clear SRE,ESE
yes
Clear PPE
yes
Clear EVENt parts of the reg-
isters
yes
yes
Clear ENABle parts of all
OPERation-and QUESTiona-
ble registers,
Fill ENABle parts of all other
registers with "1".
yes
yes
Fill PTRansition parts with "1"
Clear NTRansition parts
yes
yes
Clear error queue
yes
yes
yes
Clear output buffer
yes
yes
yes
1)
1)
1)
Clear command processing
and input buffer
yes
yes
yes
1) Every command being the first in a command line, i.e. immediately following a
<PROGRAM MESSAGE TERMINATOR>
clears the output buffer.
10.5
VNA Command Reference
This chapter describes the SCPI commands implemented by the VNA mode of the
R&S
ZNL/ZNLE.
Compatibility with R&S ZVB and older instruments
The SCPI command set for the R&S
ZNL/ZNLE vector network analyzer has been
designed for compatibility with network analyzers R&S ZVA and R&S ZVB. A special
subset of commands has been implemented for compatibility with older analyzers of
the R&S ZVR family. These commands are listed in
Chapter 10.5.3, "R&S ZVR/ZVABT
If you want to make full use of the R&S
ZNL/ZNLE features but do not need R&S ZVR
compatibility, you should use the commands listed in
VNA Command Reference