Command Reference
R&S
®
ZNL/ZNLE
844
User Manual 1178.5966.02 ─ 07
<TestPort2>
Second calibrated port number.
<TestPort3 or
AdapterCalKit>
Either the third calibrated port number, or - for Adapter Removal
calibrations only - the file path (relative to the current directory)
where the analyzer shall store the adapter characterization
obtained during the calibration.
The latter is optional. If a file path is provided the characteriza-
tion file (in Touchstone s2p format) is generated when the cor-
rection terms are finally calculated (
]CORRection:COLLect:SAVE:SELected[:DUMMy]
).
<TestPorts>
More ports to be calibrated.
Example:
CORRection:COLLect:METHod:DEFine 'Test',RSHort,
1
Select a one-port normalization at port 1 with a short standard
as calibration type.
Manual operation:
The supported calibration types are listed below.
Parameter
Type ("Calibration Presetting" dialog)
REFL | RSHort
Refl Norm Open | Refl Norm Short
FOPort
Full One Port ("Refl OSM")
FTRans | RTRans | FRTRans
Trans Norm - Forward | Trans Norm - Reverse | Trans Norm Both
"Forward" ("Reverse") means that the port with the smaller (larger)
port number serves a a drive port.
OPTPort
One Path Two Ports
TOSM
TOSM
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault
Generates a set of default system error correction data for the selected ports and cali-
bration type. The default data set corresponds to a test setup which does not introduce
any systematic errors; none of the measurement results acquired previously
(
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected
account.
Tip:
The main purpose of the default correction data set is to provide a dummy system
error correction which you can replace with your own, external correction data. You
may have acquired the external data in a previous session or even on an other instru-
ment. If you want to use the external correction data on the analyzer, simply generate
the default data set corresponding to your port configuration and calibration type and
overwrite the default data. For details refer to the program example below.
Suffix:
<Ch>
.
Channel number of the calibrated channel.
VNA Command Reference