VNA GUI Reference
R&S
®
ZNL/ZNLE
375
User Manual 1178.5966.02 ─ 07
●
"b1 Source Port 2" is the wave received at physical port 1. In a standard S-parame-
ter measurement, this wave is transmitted at port 2 of the DUT (reverse measure-
ment).
●
"b2 Source Port 2" is the wave received at physical port 2. In a standard S-parame-
ter measurement, this wave is fed to the output port (port 2) of the DUT (reverse
measurement).
Tip:
In the trace list, the source port is indicated in brackets. For example, "a1(P1)"
denotes the wave a
1
with source port 1.
The analyzer can also measure arbitrary wave quantities for other source ports; see
Chapter 9.7.3.2, "More Wave Quantities Dialog"
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "A1" | ...
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "A1" | ...
9.7.3.2
More Wave Quantities Dialog
The "More Wave Quantities" dialog provides arbitrary wave quantities with arbitrary
source ports as measured quantities. All wave quantities can be calculated with differ-
ent detector settings.
Access:
[Meas] > "Wave" > "More Wave Quantities..."
Background information
Refer to the following sections:
●
Chapter 8.3.4.1, "Wave Quantities"
●
The notation for wave quantities follows the usual scheme of the vector network ana-
lyzer:
●
The a-waves are the outgoing/transmitted waves at the analyzer's test ports.
●
The b-waves are the incoming/measured waves.
Meas Softtool