VNA GUI Reference
R&S
®
ZNL/ZNLE
484
User Manual 1178.5966.02 ─ 07
This panel can also be activated by selecting the
softtool tab.
Refer to its description of this softtool tab for background information, parameters and
additional remote commands.
Active
The checkboxes in the "Active" column activate or deactivate the configured differential
match embedding (i.e. adds or removes it to/from the calculation flow) without chang-
ing its parameters.
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:
EMBedding<LogPt>[:STATe]
File Name 1
The ellipsis button in the "File Name 1" column is enabled if a 2-port data
selected.
When loading the touchstone file (
*.s2p
), the analyzer by default assumes odd ports
left (VNA side), even ports right (DUT side). However, it is also possible to
"Swap
Gates"
, instructing the analyzer to reinterpret the loaded S-parameters (e.g. S
12
-->
S
21
).
Note:
The loaded S-parameter traces are stored in the active channel setup. Persisted
channel setups contain the full embedding and deembedding data so that they can be
transferred to other instruments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:DIFFerential:EMBedding<LogPt>
9.14.2
Offset Tab
Defines length offset parameters for each port.
Use the complementary dock widget to activate or deactivate length/loss compensation
for selected ports (see
Chapter 9.14.1.2, "Offset Panel"
The marker function
function overwrites the offset parameters.