VNA GUI Reference
R&S
®
ZNL/ZNLE
429
User Manual 1178.5966.02 ─ 07
Remote command:
CALCulate<Chn>:MARKer<Mk>[:STATe]
Search Config
Allows you to select the
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
Threshold Settings
Defines a threshold for (single) peak searches and activates it.
If "Active", only maxima above and minima below the configured "Threshold" are con-
sidered.
The unit is derived from the active trace format and cannot be modified. Changing the
trace format resets the "Threshold" to a format-specific default value.
Remote command:
CALCulate<Chn>:MARKer<Mk>:THReshold
CALCulate<Chn>:MARKer<Mk>:THReshold:STATe
Excursion Settings
Defines a minimum excursion for (single) peak searches and activates it.
If "Active", only peaks with an excursion above the configured "Excursion" value are
considered. By definition, the excursion of a peak is the smaller of the absolute differ-
ences in measured values from the adjoining peaks of opposite polarity.
The unit is derived from the active trace format and cannot be modified. Changing the
trace format resets the "Excursion" to a format-specific default value.
Remote command:
CALCulate<Chn>:MARKer<Mk>:EXCursion
CALCulate<Chn>:MARKer<Mk>:EXCursion:STATe
9.11.2
Multiple Peak Tab
"Multiple Peak" search allows you to find multiple local minima/maxima at once.
Background information
"Basic Marker Search Functions"
Marker Search Softtool
www.allice.de
Allice Messtechnik GmbH