Command Reference
R&S
®
ZNL/ZNLE
662
User Manual 1178.5966.02 ─ 07
CALCulate<Chn>:RIPPle:STATe
<Boolean>
Switches the ripple limit check for the active trace on or off.
Tip:
CALCulate<Chn>:RIPPle:SEGMent<Seg>[:STATe]
to switch the limit
check for a single ripple limit range on or off.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF
–
ripple limit check on or off.
*RST:
OFF
Example:
Manual operation:
CALCulate<Chn>:RIPPle:STATe:AREA
<HorizontalPos>, <VerticalPos>
Moves the ripple test info field for the active trace <Chn> to one of nine predefined
positions in the active diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<HorizontalPos>
LEFT | MID | RIGHt
Horizontal position
<VerticalPos>
TOP | MID | BOTTom
Vertical position
Example:
Manual operation:
CALCulate:SMOothing...
The
CALCulate:SMOothing...
commands provide the settings for trace smoothing.
CALCulate<Chn>:SMOothing[:STATe]
..............................................................................662
CALCulate<Chn>:SMOothing:APERture
.......................................................................... 663
CALCulate<Chn>:SMOothing[:STATe]
<Boolean>
Enables or disables smoothing for trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - smoothing on or off.
*RST:
OFF
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH