Command Reference
R&S
®
ZNL/ZNLE
834
User Manual 1178.5966.02 ─ 07
RSHort
–
"Refl Norm Short" calibration
UTRans
–
full n-port (UOSM) calibration
<Characterization>
Location of the characterization (cal kit file) to be used for the
automatic calibration (string parameter):
–
The empty string (
' '
) refers to the factory calibration of (and
stored on) the active calibration unit.
–
A VNA cal kit file name
*.calkit
without path refers to a
specific cal kit file stored in the internal memory of the active cal-
ibration unit.
–
A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
–
A VNA cal kit file name
*.calkit
with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
<TestPort1>
Test port numbers. For an n-port automatic calibration, n arbi-
trary (not necessarily consecutive) port numbers must be speci-
fied.
<TestPort2>
Second test port number of the analyzer
Example:
CORR:COLL:AUTO:TYPE FNPort, '', 1, 2, 4
Perform an automatic full 3-port calibration at test ports 1, 2,
and 4 using the calibration unit's default calibration kit file and
arbitrary test ports of the cal unit.
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:CKIT:INSTall
<CalKitFile>, <Gender>[,
<Ports>]
Allows you to load cal kit data by gender:
●
From the given file
●
To the given ports or all ports
on page 766 to get information about cal kit files.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<CalKitFile>
Path to the cal kit file, either absolute or relative to the current
directory (see
)
<Gender>
MALE | FEMale | NGENder
<Ports>
Either a comma-separated list of port numbers or ALL (optional)
Usage:
Setting only
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH