Command Reference
R&S
®
ZNL/ZNLE
858
User Manual 1178.5966.02 ─ 07
[SENSe<Ch>:]COUPle
<Order>
Determines the order of partial measurements and sweeps.
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the sweep mode applies
to all channels in the active channel setup.
Parameters:
<Order>
ALL | AUTO | NONE
AUTO - optimized display update: fast sweeps are performed in
alternated mode, slower sweeps in chopped mode
ALL - chopped sweep mode, complete all partial measurements
before proceeding to the next sweep point
NONE - alternated sweep mode on, reverse the order of partial
measurements and sweeps
*RST:
NONE
Example:
COUP NONE
Activate the alternated sweep mode.
TRIG:LINK 'PPO'
Set the triggered measurement sequence equal to one partial
measurement. Each trigger event starts one partial measure-
ment for all sweep points.
Manual operation:
See
[SENSe:]FREQuency...
The
[SENSe:]...
commands set frequency-related parameters, especially the mea-
surement and display ranges for the different sweep types. The frequency ranges for
the different instrument models are listed below; for more details refer to the data
sheet.
Table 10-17: Frequency ranges of R&S
ZNL analyzers
Frequency settings
Start, Stop
Center
Span
R&S
ZNL/ZNLE3
5 kHz to 3.0 GHz
>5 kHz to <3.0 GHz
1 Hz to 2.999995 GHz
R&S
ZNL/ZNLE6
5 kHz to 6 GHz
>5 kHz to <6 GHz
1 Hz to 5.999995 GHz
Table 10-18: Frequency ranges of R&S
ZNLE analyzers
Frequency settings
Start, Stop
Center
Span
R&S
ZNL/ZNLE3
1 MHz to 3.0 GHz
>1 MHz to <3.0 GHz
1 Hz to 2.999 GHz
R&S
ZNL/ZNLE3 with
100 kHz to 3.0 GHz
>100 kHz to <3.0 GHz
1 Hz to 2.9999 GHz
R&S
ZNL/ZNLE6
1MHz to 6 GHz
>1 MHz to <6 GHz
1 Hz to 5.999 GHz
R&S
ZNL/ZNLE6 with
100 kHz to 6.0 GHz
>100 kHz to <6.0 GHz
1 Hz to 5.9999 GHz
VNA Command Reference