Getting Started
R&S
®
ZNL/ZNLE
54
User Manual 1178.5966.02 ─ 07
5.2.2.10
Aux. Port
A 25-pole D-Sub connector used as an input and output for low-voltage TTL control
signals. Can either be used as VNA User Port (max. 3.3
V) or SA Aux. Port (max. 5
V).
At the FW GUI this can be set via [Setup] > "System Config" > "Add Interfaces" > "Port
Configuration".
13
1
14
25
This connector is provided by the "Additional Interfaces" option R&S
FPL1-B5.
For details see
This option is not available for the R&S
ZNLE.
Short-circuit hazard
Always observe the designated pin assignment. A short-circuit can damage the port.
5.2.2.11
IF/Video Output
The female BNC connector can be used for various outputs in the Spectrum applica-
tion:
●
Intermediate frequency (IF) output of approximately 20 MHz
●
Video output (1V)
Which output is provided is defined in the software ("Overview" > "Output" ).
●
This connector is provided by the "Additional Interfaces" option R&S
FPL1-B5.
This option is not available for the R&S
ZNLE.
●
The IF/Video Output interface is only functional in Spectrum mode (option
R&S
ZNL3|6-B1)
●
Interruptions in analog output may occur if you switch between a VNA and a spec-
trum application, due to the separate hardware. In particular, if you run a sequence
of measurements including both measurement types, switching between both
modes can interrupt the signal from the output connector.
For details, see the R&S
ZNL/ZNLE Spectrum Mode User Manual.