VNA GUI Reference
R&S
®
ZNL/ZNLE
400
User Manual 1178.5966.02 ─ 07
If enabled, an additional info field is displayed in the diagram, indicating the "PASS" or
"FAIL" state. Limit violations are marked with a colored square. An acoustic signal
(
) and a TTL signal indicating pass or fail can be generated in addition.
Note:
●
Limit check and display of limit lines are independent of each other:
–
The limit lines can be displayed, no matter if the limit check is enabled.
–
If "Limit Check" is enabled, the limits are checked, no matter if the limit lines
are displayed.
–
The limit check can even be enabled, if no limit lines are defined. In this case,
the info field displays "No limit defined!" and the limit check always passes.
●
Limit lines are defined for a particular trace format. However, the limit check is per-
formed irrespective of the current trace format. The info field indicates the correct
"PASS"/"FAIL" state and limit violations are visualized on the trace (if any).
Figure 9-9: Limit line violations in complex trace formats
Remote command:
CALCulate<Chn>:LIMit:STATe
CALCulate<Chn>:LIMit:LOWer:STATe
CALCulate<Chn>:LIMit:UPPer:STATe
CALCulate<Chn>:LIMit:FAIL?
CALCulate:LIMit:FAIL:ALL?
CALCulate<Chn>:LIMit:STATe:AREA
Limit Fail Beep
Activates or deactivates the fail beep. The fail beep is a low-tone acoustic signal that is
generated each time the analyzer detects an exceeded limit. No fail beep can be gen-
erated if the limit check is switched off.
Note:
The R&S
ZNL/ZNLE does not have a built-in audio device and loudspeaker. To
hear these sounds, connect a USB audio device to the instrument or operate it via
remote desktop.
Display Lines Softtool