Command Reference
R&S
®
ZNL/ZNLE
681
User Manual 1178.5966.02 ─ 07
An execution error is raised, if the calculated number of points is higher than the maxi-
mum number of sweep points the firmware allows (100001). To avoid this, reduce the
frequency span Δf (see
on page 883) and/or the
stop frequency d
stop
(see
CALCulate<Chn>:TRANsform:DTFault:STOP
2.6 · d
stop
· Δf / (v · c
0
) ≤ 100001
,
where
v
denotes the velocity factor of the measured cable (see
).
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
*RST; :CALCulate1:TRANsform:DTFault:STATe ON
makes the active trace of channel 1 a DtF trace.
SENSe1:SWEep:POINts?
returns the default number of sweep points (201).
:CALCulate1:TRANsform:DTFault:STARt 0; STOP 100
widens the DtF distance window to 100m.
CALCulate1:TRANsform:DTFault:POINts
starts the "Auto Number of Points calculation". Now
SENSe1:SWEep:POINts?
returns an increased number of sweep points.
Usage:
Event
Options:
R&S
ZNL-K3
Manual operation:
See
CALCulate<Chn>:TRANsform:DTFault:STARt
<Start>
CALCulate<Chn>:TRANsform:DTFault:STOP
<Stop>
Defines the window of the Distance to Fault measurement using its start and stop dis-
tance.
See also
CALCulate<Chn>:TRANsform:DTFault:CENTer
CALCulate<Chn>:TRANsform:DTFault:SPAN
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Start>/<Stop>
Start/stop distance.
Default unit: m
Example:
:CALCulate1:TRANsform:DTFault:STARt 1; STOP 3
Defines the DtF distance window of the active trace of channel 1
as [1m, 3m].
Options:
R&S
ZNL-K3
Manual operation:
See
"Start Distance / Stop Distance"
VNA Command Reference