Command Reference
R&S
®
ZNL/ZNLE
857
User Manual 1178.5966.02 ─ 07
Example:
*RST; CORR:OFFS:DFC?
Reset the instrument and query whether the analyzer uses direct
fixture compensation data at port 1. The response is 0.
Usage:
Query only
Manual operation:
See
[SENSe<Ch>:]CORRection:STIMulus?
Queries the stimulus values of the active calibration. A calibration must be selected
before the command is executed; see example.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example:
*RST; :CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORR:STIM?
Query the stimulus frequencies. The response contains 201 fre-
quency values.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE:SEL
Calculate the system error correction data and apply it to the
active channel.
Usage:
Query only
Manual operation:
See
[SENSe<Chn>:]CORRection:SSTate?
Returns the system error correction state label of the active trace in channel
<Chn>
.
The response is a string variable containing the calibration state label in the trace list
('CAL', 'CAI', 'CAL OFF' ..; see
Chapter 8.5.4, "Calibration Labels"
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
See
Usage:
Query only
Manual operation:
See
[SENSe:]COUPle...
The
[SENSe:]COUPle...
commands select the sweep mode.
VNA Command Reference