Manual Trigger
Generates the trigger event for "Manual" trigger mode and is disabled unless this mode
is active.
Remote command:
TRIGger<Ch>[:SEQuence]:SOURce MANual
*TRG
Sequence
Selects the measurement cycle or sequence of actions to be triggered in "External" or
"Manual" mode.
●
"Sweep" means that each trigger event starts an entire sweep, according to the
current sweep configuration.
●
"Point" means that each trigger event starts the measurement at the next sweep
point.
●
"Partial Measurement" means that each trigger event starts the next partial mea-
surement at the current or at the next sweep point. If every sweep point only
requires a single partial measurement, this option is equivalent to "Point".
See also
Chapter 8.1.4.1, "Partial Measurements and Driving Mode"
●
"Segment" means that each trigger event starts the next sweep segment within the
current sweep. If a sweep type other than
is active, this option is equiv-
alent to "Sweep".
Relation with other sweep settings
Some sweep settings are logically incompatible with a particular selection of the trig-
gered measurement sequence:
●
sweep is performed, the sequence is always a sweep.
●
only makes sense if the triggered measurement
sequence comprises more than one sweep point. If "Point" or "Partial Measure-
ment" is selected, "Alternated" mode is switched off and vice versa.
Note:
The trigger events must be adjusted to the triggered measurement sequence. If
the analyzer receives a trigger event while the last sequence is still running, the
R&S
ZNL/ZNLE skips the trigger event and generates a message.
Remote command: