Command Reference
R&S
®
ZNL/ZNLE
877
User Manual 1178.5966.02 ─ 07
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]
<IntSourcePow>
Defines the power of the internal signal source in sweep segment no.
<Seg>
. At the
same time, the command activates separate power control in all sweep segments
(
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]:CONTrol
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Parameters:
<IntSourcePow>
Internal source power
Range:
-300 dBm to +200 dBm. The usable power range is
frequency-dependent; refer to the data sheet.
*RST:
-10 dBm
Default unit: dBm
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings and thus -10 dBm internal source power.
SEGM:POW -20
Decrease the power to -20 dBm.
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]:CONTrol
<Boolean>
Selects common or independent internal source "Power" settings for the sweep seg-
ments.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. This suffix is ignored; the setting con-
trols the whole segmented sweep.
Parameters:
<Boolean>
ON
–
use independent power settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]
OFF
–
reset the power in all sweep segments to the power for
unsegmented sweeps, defined via
POWer<PhyPt>[:LEVel][:IMMediate][:AMPLitude]
will not restore the previous values.
The parameter is automatically switched to
ON
when a segment
power is entered using
*RST:
OFF