Command Reference
R&S
®
ZNL/ZNLE
826
User Manual 1178.5966.02 ─ 07
Example:
Suppose there are 3 channels in the current channel setup and
you want to calibrate channels 1 and 2. For channel 1 you want
to perform a Reflection Normalization (Short) on ports 1 and 2,
for channel 2 an OSM calibration on port 2. This requires mea-
suring standards as shown in the table below. Proceed as fol-
lows:
:CORRection:COLLect:CHANnels:MCTYpes ON
Activate the MCTYPes mode.
:SENSe1:CORRection:COLLect:METHod:DEFine
'Parser Test SFK Ch1', RSHort, 1, 2
:SENSe2:CORRection:COLLect:METHod:DEFine
'Parser Test SFK Ch2', FOPort, 2
Declare the calibrations to be performed, then connect the Short
standard to port 1
:SENSe1:CORRection:COLLect:ACQuire:SELected
SHORt, 1
Acquire calibration data for channel 1, then connect the Open
standard to port 2
:SENSe2:CORRection:COLLect:ACQuire:SELected
OPEN, 2
Acquire calibration data for channel 2, then connect the Short
standard to port 2
:SENSe1:CORRection:COLLect:ACQuire:SELected
SHORt, 2
:SENSe2:CORRection:COLLect:ACQuire:SELected
SHORt, 2
Acquire calibration data for channel 1 and 2, then connect the
Match standard to port 2
:SENSe2:CORRection:COLLect:ACQuire:SELected
MATCh, 2
Acquire calibration data for channel 2
:SENSe1:CORRection:COLLect:SAVE:SELected
Complete all calibrations. This command is required only once
and has to address one of the calibrated channels (channel 1 or
2 in this example)
:CORRection:COLLect:CHANnels:MCTYpes OFF
Deactivate the MCTYPes mode. Recommended for compatibility
with manual operation!
Port 1
Port 2
Channel 1
Reflection Normalization(Short)
for ports 1 and 2
Short
Short
Channel 2
OSM for port 1
-
Open, Short, Match
VNA Command Reference