Command Reference
R&S
®
ZNL/ZNLE
886
User Manual 1178.5966.02 ─ 07
[SENSe<Ch>:]SWEep:TIME:AUTO
<Boolean>
When enabled, the (minimum) sweep time is calculated internally using the other chan-
nel settings and zero delay (
Note:
The automatically calculated sweep duration is ignored for the sweep types
).
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
ON | OFF - turn the automatic calculation of the sweep time on
or off.
OFF is also set if the sweep duration or delay is set explicitly
using
*RST:
ON
Example:
SWE:TIME 1
Set a total sweep time of 1 s.
SWE:TIME:AUTO?
A query returns the value
1
.
Manual operation:
See
[SENSe<Ch>:]SWEep:TYPE
<Format>
Selects the sweep type, i.e. the sweep variable (frequency/time) and the position of the
sweep points across the sweep range.
Suffix:
<Ch>
.
Channel number.
Parameters:
<Format>
LINear | LOGarithmic | CW | POINt | SEGMent
LINear - linear frequency sweep at constant source power
(
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate][:
(
[SENSe<Ch>:]FREQuency:...
) is swept in equidistant steps
over the frequency range. In a Cartesian diagram, the x-axis is a
linear frequency axis.
LOGarithmic - logarithmic frequency sweep. The frequency is
swept in equidistant steps on a logarithmic scale. In a Cartesian
diagram, the x-axis is a logarithmic frequency axis.
CW - time sweep. The measurement is performed at constant
frequency (
SOURce<Ch>:FREQuency<PhyPt>:FIXed
SOURce<Ch>:POWer<PhyPt>[:LEVel][:
) and repeated over a specified
period of time at constant time intervals.
POINt - CW mode sweep, time sweep triggered according to the
current trigger settings.