Command Reference
R&S
®
ZNL/ZNLE
884
User Manual 1178.5966.02 ─ 07
Parameters:
<SweepPoint>
Number of points per sweep
Range:
see above
*RST:
201
Example:
*RST
Reset the instrument, activating a frequency sweep with 201
sweep points.
SWE:TIME?
Query total sweep time.
SWE:POIN 2010
Multiply the (default) number of points by 10.
SWE:TIME?
Query total sweep time again. The analyzer estimates a sweep
time that is also multiplied by 10.
Manual operation:
See
[SENSe<Ch>:]SWEep:STEP
<StepSize>
Sets the distance between two consecutive sweep points in a (non-segmented) linear
frequency sweep.
Suffix:
<Ch>
.
Channel number
Parameters:
<StepSize>
Frequency step size.
The step size is equal to the current sweep spandivided by the
number of sweep points minus one (see
respectively). See also the description of manual control and the
program example below.
Range:
Depends on the current sweep span and the maxi-
mum number of sweep points: "Span Frequency" /
(max. "Number of Points" – 1) ≤ "Freq Step Size" ≤
"Span Frequency"
*RST:
Depends on the analyzer model. The default step
size is equal to the default sweep span of the ana-
lyzer divided by the default number of sweep points
minus one.
Default unit: Hz