DIAGnostic:DUMP:SIZE
<DumpSize>
Determines the level of detail and hence the size of the dump files created in case of
firmware exceptions.
Parameters:
<DumpSize>
NONE | MINI | NORMal | LARGe | FULL
Either disables dump file creation (NONE) or determines the
level of detail.
DIAGnostic:SERVice:RFPower
<Boolean>
Turns the internal source power at all ports and the power of all external generators on
or off. This command is equivalent to
Parameters:
<Boolean>
ON | OFF - switch the power on or off.
*RST:
ON
Example:
DIAG:SERV:RFP OFF
Turn off the RF source power.
10.5.2.5
DISPlay Commands
The
DISPlay...
commands control the selection and presentation of graphical and
trace information on the screen.
Trace display
Traces are identified by a string parameter defining the trace name (e.g.
CALCulate<Ch>:PARameter:SELect
<TraceName>
). In the
DISPlay...
subsys-
tem, traces are assigned to diagrams (
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:
<TraceName>
). While this assignment is valid, the trace is identified by the
numeric suffix
<Wnd>
, and the trace name is not needed.
Units for DISPlay... commands
The
DISPlay...
subsystem contains commands to define particular points in the dia-
gram, e.g. to set the scale or a reference value. Some settings require a numeric value
and a physical unit, depending on the related parameter type. The following table lists
the physical units accepted by the analyzer.
Parameter type
Physical unit
Power
DBM, DB, DBW, W, MW, UW, NW, PW
Voltage
V, MV, UV, NV, PV, DBV, DBMV, DBUV
Phase
DEG, KDEG, MDEG, UDEG, NDEG, PDEG
Group delay
S, MS, US, NS, PS