VNA GUI Reference
R&S
®
ZNL/ZNLE
443
User Manual 1178.5966.02 ─ 07
Note:
Transmission normalization and "One Path Two Ports" calibration types require
two-port (Through) characterization data for the cal unit. These two-port characteriza-
tions can be unavailable in the factory characterizations of some older calibration units
or in a user characterization. If a tooltip indicates missing two-port characterization
data, simply perform a new characterization of your cal unit. In the first dialog of the
"Characterization" wizard, select "Take All OSM and Through" to make sure that the
necessary two-port data is acquired. See also
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
Next
Proceeds to
. "Next" is unavailable (and a warning is displayed) if
the following happens:
●
The selected characterization data do not cover all the ports to be calibrated.
●
The selected characterization data do not contain all standards needed for the
selected calibration type.
Step 2: Connections
Defines the port assignments between the R&S
ZNL/ZNLE and the calibration unit.
Figure 9-12: Calibration Unit wizard, step 2: Connections
Port Assignment (manual)
The "Connections" step allows you to configure the assignment between VNA ports
and calibration unit ports manually.
The test port connectors are automatically set according to the connector type of the
selected calibration unit port.