VNA GUI Reference
R&S
®
ZNL/ZNLE
408
User Manual 1178.5966.02 ─ 07
Show Results All Traces
Defines the visibility of ripple info fields in the active recall set.
●
If disabled, only the ripple info field of the active trace is displayed (in case it has
ripple check enabled).
●
If enabled, ripple info fields are displayed for all traces with ripple check enabled.
Remote command:
CALCulate:RIPPle:DISPlay:RESult:ALL[:STATe]
Clear Test
Resets the limit check results.
Remote command:
Global Check
TTL1 Pass / TTL2 Pass
9.9.5.2
Define Ripple Test Dialog
The "Define Ripple Test" dialog defines the ripple limits for the active trace on a range-
by-range basis. A separate ripple limit can be assigned to each range.
Access:
[Display
Lines] > "Ripple Test" > "Def. Ripple Test..."
Defining ripple limits with minimum effort
Choose one of the following methods to create and handle ripple limit ranges effi-
ciently:
●
To configure a limit test with only a few ranges, use "Add" and edit each range in
the table individually.
●
Use the "Align All" button to create non-overlapping, contiguous ranges of equal
width.
●
Save your ripple ranges to a file so you can reuse or modify them in later sessions
("Save Ripple Test..., Recall Ripple Test...").
Background information
Display Lines Softtool
www.allice.de
Allice Messtechnik GmbH