VNA Concepts and Features
R&S
®
ZNL/ZNLE
284
User Manual 1178.5966.02 ─ 07
8.7.5
Extended Power Range
Option R&S
ZNL3|6-B22
For the R&S
ZNL3|6, hardware option R&S
ZNL3|6-B22 reduces the minimum receive
power down to -40 dBm.
This option is not available for the R&S
ZNLE.
8.7.6
Receiver Step Attenuators
Options R&S
ZNL3|6-B31 and R&S
ZNL3|6-B32
Receiver step attenuators are used to adjust the received signal levels to the input
level range of the analyzer to avoid damage to the instrument, e.g. if the DUT is a
power amplifier. The attenuation values can be set between 0 dB and 30 dB (in 10 dB
steps); see
on page 291. The following step attenuators are availa-
ble:
Table 8-13: R&S
ZNL
Analyzer type
Port 1
Port 2
R&S
ZNL3
R&S
ZNL3-B31
R&S
ZNL3-B32
R&S
ZNL6
R&S
ZNL6-B31
R&S
ZNL6-B32
●
Receiver step attenuators are not available for the R&S
ZNLE.
●
Receiver step attenuators for the R&S
ZNL can be retrofitted at Rohde
&
Schwarz
service.
8.7.7
Lower Start Frequency for R&S
ZNLE
Option R&S
ZNLE-B100
Hardware option R&S
ZNLE-B100 reduces the lower frequency bound of a R&S
ZNLE
from 1
MHz down to 100
kHz.
●
The frequency range of a R&S
ZNL always starts at 5
kHz
●
Option R&S
ZNLE-B100 can be retrofitted at Rohde
&
Schwarz service.
Optional Extensions and Accessories