Command Reference
R&S
®
ZNL/ZNLE
759
User Manual 1178.5966.02 ─ 07
Parameters:
<RelFontSize>
Relative font size
Range:
80 % to 170 %
*RST:
100 %
Default unit: percent
Example:
*RST; :DISP:RFS 80
Use smaller fonts to gain more space for the traces in the dia-
gram.
Manual operation:
10.5.2.6
FORMat Commands
The
FORMAT...
commands select a data format for transferring numeric data (includ-
ing arrays) from and to the analyzer.
FORMat[:DATA]
<TransferFormat>[, <Length>]
Selects the format for numeric data transferred to and from the analyzer.
Note:
The format setting is only valid for commands and queries whose description
states that the response is formatted as described by
FORMat[:DATA]
. In particular, it
affects trace data transferred by means of the commands in the
TRACe:...
system.
Parameters:
<TransferFormat>
ASCii | REAL
ASCii - numeric data is transferred as ASCII bytes. The numbers
are separated by commas as specified in IEEE 488.2.
REAL - Data is transferred in a definite length block as IEEE
floating point numbers of the specified <Length>. See
ter 10.4.1, "Block Data Format"
Note:
If binary data is transferred to the analyzer, the receive
terminator should be set to EOI (
) to avoid inadvertent inter-
ruption of the data transfer.
<Length>
The optional <Length> parameter is needed for REAL format
only. It defines the length of the floating point numbers in bits.
Valid values are 32 and 64.
*RST:
ASCII. The default length of REAL data is 32 bits
(single precision).
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH