Connector and calibration kit names must be entered as string parameters. The strings
contain the connector and cal kit names used in the
Calibration Kits
dialog; a
Ω
in the name must be replaced by 'Ohm', e.g.:
–
'NewKit1'
denotes the user-defined calibration kit "NewKit1".
–
'N 50 Ohm Ideal Kit'
denotes the "N 50
Ω
Ideal Kit".
–
'ZV-Z21 typical'
denotes the cal kit "ZV-Z21 typical".
MMEMory:LOAD:CKIT:SDATa:WLABel
<ConnectorType>, <CalKitName>,
<KitLabel>, <StandardType>, <StandardLabel>, <TouchstoneFile>[, <FirstPort>,
<SecondPort>]
Loads characterization data from the given Touchstone file; similar to existing com-
mand
but supports cal kit addressing
by label
.
Setting parameters:
<ConnectorType>
String parameter containing the name of the connector type.
<CalKitName>
String parameter containing the name of a calibration kit availa-
ble on the analyzer.
<KitLabel>
String parameter containing the label of a calibration kit available
on the analyzer. An empty string means that no label is defined.
<StandardType>
MMTHrough | MFTHrough | FFTHrough | MMLine | MMLine1 |
MMLine2 | MMLine3 | MFLine | MFLine1 | MFLine2 | MFLine3 |
FFLine | FFLine1 | FFLine2 | FFLine3 | MMATten | MFATten |
FFATten | MMSNetwork | MFSNetwork | FFSNetwork | MOPen |
FOPen | MSHort | FSHort | MOSHort | MOSHort1 | MOSHort2 |
MOSHort3 | FOSHort | FOSHort1 | FOSHort2 | FOSHort3 |
MREFlect | FREFlect | MMTCh | FMTCh | MSMatch | FSMatch
Standard types; for a description refer to table
<StandardLabel>
String parameter addressing a particular calibration standard by
its label. An empty string means that no label is defined.
<TouchstoneFile>
String parameter to specify the name and directory of the Touch-
stone file to be loaded. A
*.s1p
file must be used for one-port
standards, a
*.s2p
file for two-port standards.
If no path is specified the analyzer searches the current direc-
tory, to be queried with
<FirstPort>
First port number (sufficient for one-port standards). If the port
numbers are omitted, the cal kit data is valid for all ports.
<SecondPort>
Second port number, for two-port standards. If the port numbers
are omitted, the cal kit data is valid for all ports.