![Rohde & Schwarz 1323.0012.03 User Manual Download Page 436](http://html1.mh-extra.com/html/rohde-and-schwarz/1323-0012-03/1323-0012-03_user-manual_1477777436.webp)
VNA GUI Reference
R&S
®
ZNL/ZNLE
436
User Manual 1178.5966.02 ─ 07
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:BWIDth:MODE BPRMarker
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute BFILter
Bandstop Ref to Max
Activates the search for a bandstop region on the active trace and activates
A bandstop region is the lowest peak (local minimum) in the search range, whose level
is at least <Bandwidth> below the maximum (passband value).
If "Bandstop Ref to Max" is selected, the analyzer uses (or creates) the four markers
"M1" to "M4" to locate the
bandstop region
.
●
"M1" indicates the minimum of the peak ("Min").
●
"M2" indicates the "Lower Edge" of the bandstop, i.e. the point on the left edge of
the peak where the trace value is equal to the maximum in the search range (pass-
band value) minus the specified
.
●
"M3" indicates the "Upper Edge" of the bandstop, i.e. the point on the right edge of
the peak where the trace value is equal to the maximum in the search range (pass-
band value) minus the specified "Bandwidth".
●
"M4" indicates the center of the peak. Depending on a system setting, the center is
either calculated as the geometric or the arithmetic mean of the "Lower Edge" and
"Upper Edge" positions (see
"Geometric Calculation of Bandfilter Center "
The search results are displayed in the movable "Bandfilter" info field.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:BWIDth:MODE BSTop
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute BFILter
Result Off
Hides the movable info field with the results of a bandpass or a bandstop search and
disables
. The info field is displayed again (and tracking re-enabled) when a
new "Bandfilter" search is performed.
Remote command:
CALCulate<Chn>:MARKer:SEARch:BFILter:RESult[:STATe]