VNA GUI Reference
R&S
®
ZNL/ZNLE
467
User Manual 1178.5966.02 ─ 07
9.12.3.2
Calibration Manager Dialog
The "Calibration Manager" dialog stores user correction data to the cal pool and
assigns stored correction data to channels.
See
Chapter 8.5.3, "Calibration Pool"
on page 253 for background information.
Access:
[Cal] > "Cal Manager..."
Channel State
The "Channel State" table shows all channels in the active channel setup together with
their current calibration. Channels can use either the active channel calibration (if avail-
able), a previously stored user correction data or the factory system error correction
(indicated as '--').
Remote command:
n/a
Add / Add All... / Replace / Apply / Apply to All
The buttons between the tables are used to modify the calibration pool and apply cali-
bration data sets (cal groups) to channels:
●
"Add" copies the correction data of the selected channel to the cal pool, generating
a new pool member (cal group).
●
"Add All..." copies the correction data of all channels to the cal pool, generating
new pool members (cal groups).
●
"Replace" overwrites a cal group with new correction data.
●
"Apply" assigns the selected cal group to the selected channel.
●
"Apply to All" assigns the selected cal group to all channels in the "Channel State"
table.