Command Reference
R&S
®
ZNL/ZNLE
898
User Manual 1178.5966.02 ─ 07
Setting parameters:
<Boolean>
ON - memory is shown in a separate drive.
OFF - memory is not shown.
*RST:
OFF
Example:
SYST:COMM:AKAL:MMEM ON
Show the memory of the active calibration unit.
Usage:
Setting only
Manual operation:
See
SYSTem:COMMunicate:GPIB[:SELF]:DCLear:SUPPress
<Boolean>
Suppresses/unsuppresses of Device Clear GPIB interface messages (DCL, SDC).
Parameters:
<Boolean>
SYSTem:COMMunicate:GPIB[:SELF]:INIT:WAIT
<Boolean>
Determines/queries the execution behavior of
INITiate[:IMMediate]
commands
Chapter 10.5.2.7, "INITiate Commands"
If set to ON, an automatic
*WAI
is added (see
on page 509), i.e. the commands
execute synchronously.
Parameters:
<Boolean>
*RST:
OFF
Manual operation:
See
"Remote Settings: Wait for Data after Sweep "
SYSTem:COMMunicate:GPIB[:SELF]:LPORt:ALIGn
<Boolean>
Configures/queries the logical port creation.
If set to ON, logical ports are aligned and must be set from low to high port, which was
the only possibility prior to firmware V1.90. If set to OFF, new ports can be created
freely, like in manual operation. See the example below.
Parameters:
<Boolean>
*RST:
ON
Example:
Starting with a 4-port analyzer's default logical port assignment
L1<-->P1, L2<-->P2, L3<-->P3, L4<-->P4,
the remote command
SOUR:LPOR1 1,2
by default (
ALIGn=ON
) generates aligned logical ports
L1<-->P1&P2, L2<-->P3, L3<-->P4.
With
ALIGn=OFF
, the result is
L1<-->P1&P2, L3<-->P3, L4<-->P4
and L2 does not exist.
VNA Command Reference