VNA GUI Reference
R&S
®
ZNL/ZNLE
466
User Manual 1178.5966.02 ─ 07
9.12.3.1
Related Controls on the Calibration Tab
The buttons in the lower part of the "Calibration" tab open the following dialogs:
●
"Manage Cals" – "Cal Manager...":
See
User Cal Active
Activates or deactivates the system error correction in the active channel. "User Cal
Active" is available only if a valid system error correction is available for the active
channel; see "Channel State" in
Chapter 9.12.3.2, "Calibration Manager Dialog"
Note:
A label "Cal Off" appears behind the trace list if the system error correction is
switched off; see also
Chapter 8.5.4, "Calibration Labels"
status of each channel and trace appears in the setup information ("Setup" > "Info..." >
"Setup").
Remote command:
[SENSe<Ch>:]CORRection[:STATe]
Manage Cals – Recall Last Cal Set
Loads and activates the channel setup for which the last calibration was performed. If
the last calibrated setup is already active, nothing is changed. The calibrated setups
are automatically stored in the
C:\Users\Public\Documents\Rohde-Schwarz\ZNL\Calibration\
RecallSets
directory. A message box pops up if the directory is empty, e.g. because
no calibration was performed yet.
Remote command:
n/a