Command Reference
R&S
®
ZNL/ZNLE
827
User Manual 1178.5966.02 ─ 07
[SENSe:]CORRection:COLLect:DETector
<mode for correction detector>
Selects the S-parameter detector type during manual calibration.
Parameters:
<mode for correction
detector>
NORMal | AVERage
AVERage
Average detector, recommended for channels measuring traces
using detectors other than the NORMal one
NORMal
Normal (sample) detector, recommended for all other applica-
tions
*RST:
NORMal
[SENSe:]CORRection:COLLect:FIXTure:LMParameter[:STATe]
<Boolean>
Selects an Auto Length (and Loss) calculation or a Direct Compensation.
Parameters:
<Boolean>
1 - Auto Length (and Loss), depending on the last
[SENSe<Ch>:]CORRection:COLLect:FIXTure:
LMParameter:LOSS[:STATe]
setting
0 - Direct Compensation
*RST:
1
Example:
See
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:
Manual operation:
See
[SENSe:]CORRection:COLLect:FIXTure:LMParameter:LOSS[:STATe]
<Boolean>
Selects an Auto Length or an Auto Length and Loss calculation.
Parameters:
<Boolean>
1 - Auto Length and Loss
0 - Auto Length, no loss
*RST:
1
Example:
See
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:
Manual operation:
See
"Auto Length / Auto Length and Loss"
[SENSe:]CORRection:IMEThod
<InterpolationMethod>
Selects the algorithm for the interpolation of user system error corrections.