Command Reference
R&S
®
ZNL/ZNLE
856
User Manual 1178.5966.02 ─ 07
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer. This numeric suffix is ignored; the
command affects the parameters of all ports.
Parameters:
<Boolean>
The parameter function depends on whether the command is
used as a setting command or as a query: For setting command:
ON - no effect
OFF - resets all length offsets to zero and the reference fre-
quency to 1 GHz
For query:
1 - at least one length offset parameter is different from its
default value
0 - all length offsets are zero / set to default
*RST:
OFF
Example:
*RST; :CORR:OFFS?
Reset the instrument and query whether the length offset param-
eters have been reset as well. The response is 0.
Manual operation:
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:COMPensation[:STATe]
<Boolean>
Toggles length/loss/fixture compensation for physical port <PhyPt> ON/OFF.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Parameters:
<Boolean>
ON (1): compensation active
OFF (0): compensation inactive
Manual operation:
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:DFComp[:STATe]?
Returns whether a direct fixture compensation has been carried out at port no.
<PhyPt>
. A direct fixture compensation resets the offset parameters to zero, the ana-
lyzer uses calculated transmission factors instead.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer.
Return values:
<Boolean>
1 - direct fixture compensation data used
0 - no direct fixture compensation data used