![Rohde & Schwarz 1323.0012.03 User Manual Download Page 498](http://html1.mh-extra.com/html/rohde-and-schwarz/1323-0012-03/1323-0012-03_user-manual_1477777498.webp)
VNA GUI Reference
R&S
®
ZNL/ZNLE
498
User Manual 1178.5966.02 ─ 07
If the "Fixture Simulator" is disabled for the related channel (see
on page 471), this tab is inactive, i.e. all controls except the "Overview" button are
grayed out.
Overview
Type
Switches between "Deembedding" and "Embedding" network definition.
Logical Port
Logical analyzer port, as defined in the "Balanced Ports" configuration. The transfor-
mation networks are defined such that the physical analyzer test ports are connected
to the left of the circuit; the DUT ports are on the right side.
Remote command:
The
<LogPt>
numeric suffix in the embedding/deembedding commands identifies the
logical port; see e.g.
CALCulate<Ch>:TRANsform:VNETworks:BALanced:
.
Network
The graphical list contains all available 4-port networks (see
The symbol selects "no network" and disables deembedding/embedding for the
selected balanced port.
Other 2-port data-subnetworks (symbols "D1" and "D2") are defined by means of s2p
files (see
The parameters of lumped elements are displayed below the graphical list.
Tip:
●
Drag and drop the network symbols in horizontal or vertical direction to switch to
the next symbol. The lumped elements and S-parameter networks ("D1", "D2") are
numbered from top to bottom.
●
Use the
Conductance in Embedding Networks
switch to change from resistances
in "Capacitor in parallel with resistor" circuit blocks to conductances and vice versa.
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
TNDefinition
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:C<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:L<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:R<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:G<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
TNDefinition
Offset Embed Softtool
www.allice.de
Allice Messtechnik GmbH