VNA GUI Reference
R&S
®
ZNL/ZNLE
485
User Manual 1178.5966.02 ─ 07
Background information
Refer to the following sections.
●
Chapter 8.6, "Offset Parameters and Embedding"
●
Chapter 8.6.1.1, "Definition of Offset Parameters "
●
Chapter 8.6.1.3, "Auto Length"
●
Chapter 8.6.1.6, "Application and Effect of Offset Parameters "
●
Chapter 8.6.1.7, "Offset Parameters for Balanced Ports"
●
Chapter 8.6.1.5, "Fixture Compensation "
9.14.2.1
Controls on the Offset Tab
The "Fixture Compensation..." button opens the
.
Overview
This button is available on all "Offset Embed" softtool tabs. It opens the
.
Port
Physical test port of the analyzer. You can define independent offset parameters for all
ports.
Remote command:
The
<PhyPt>
numeric suffix in the
[SENSe<Ch>:]CORRection:...
commands
identifies the physical port.
Delay / Electrical Length / Mech. Length
Defines the length offset at the selected port as a delay, an electrical length, or a
mechanical length. The three quantities are related by:
Offset Embed Softtool
www.allice.de