Command Reference
R&S
®
ZNL/ZNLE
709
User Manual 1178.5966.02 ─ 07
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding:TNDefinition
<CircuitModel>
Selects the circuit model for ground loop embedding.
Suffix:
<Ch>
.
Channel number.
Parameters:
<CircuitModel>
FIMPort | SL | SC | SG
Possible circuit models (character data); see
ground loop port embedding/deembedding
*RST:
FIMPort
Example:
See
CALCulate<Ch>:TRANsform:VNETworks:GLOop:
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:FSIMulator[:STATe]
<Enable>
De/activates the "Fixture Simulator" switch that allows to disable and (re-)enable the
configured deembedding, embedding, balanced ports, and port impedance settings for
the selected channel.
Suffix:
<Ch>
.
Channel number
Parameters:
<Enable>
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding:DELete
Deletes all port sets (including port pairs) previously defined for deembedding.
Suffix:
<Ch>
.
Channel number
Example:
See
CALCulate<Ch>:TRANsform:VNETworks:PPAir:
Usage:
Event
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>[:STATe]
<arg0>
Enables or disables the deembedding function for port set (or port pair) <ListId>. It is
allowed to change the deembedding network while embedding is enabled.
VNA Command Reference
www.allice.de