Command Reference
R&S
®
ZNL/ZNLE
787
User Manual 1178.5966.02 ─ 07
Example:
MMEM:LOAD:CKIT 'C:\Users\Public\Documents
\Rohde-Schwarz\Vna\Calibration\Kits\New Kit
1.calkit'
Load the previously created cal kit file
New Kit 1.calkit
from the default cal kit directory.
... :MMEM:STOR:CKIT 'New Kit 1', 'C:
\Users\Public\Documents
\Rohde-Schwarz\Vna\Calibration\Kits\New Kit
1.calkit'
Store the data for the user-defined cal kit "New Kit 1" and over-
write the cal kit file
New Kit 1.calkit
.
Usage:
Setting only
Manual operation:
See
"Import Cal Kit... / Export Cal Kit..."
MMEMory:STORe:CKIT:WLABel
<CalKitName>, <KitLabel>, <CalKitFile>
Stores the data of a calibration kit to a specified file. The calibration kit is identified by
its name and label.
Setting parameters:
<CalKitName>
Name of a user-defined calibration kit available on the analyzer.
Tip:
It is not possible to modify or store predefined or ideal kits.
<KitLabel>
Label of the calibration kit, usually its serial number.
<CalKitFile>
String parameter to specify the name and directory of the cal kit
file to be created. The file is a NWA-specific cal kit file with the
extension *.calkit.
If no path is specified the analyzer searches the current direc-
tory, to be queried with
Example:
See
[SENSe:]CORRection:CKIT:LLABel
Usage:
Setting only
MMEMory:STORe:CORRection
<Channel>, <CalGroupFile>
Copies the correction data of channel
<Channel>
to the cal pool, generating a new
correction data file (cal group). The file has the extension
*.cal
and is stored in the
C:\Users\Public\Documents\Rohde-Schwarz\ZNL\Calibration\Data
directory.
Setting parameters:
<Channel>
Channel number
<CalGroupFile>
String parameter to specify the name of the created cal group
file. There is no need to specify the directory path and file exten-
sion; the analyzer uses the default cal pool directory
C:\Users\Public\Documents\Rohde-Schwarz\ZNL
\Calibration\Data
and a
*.cal
extension.
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH