Command Reference
R&S
®
ZNL/ZNLE
624
User Manual 1178.5966.02 ─ 07
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:EXCursion
<Value>
Defines a marker-specific minimum excursion value for (single) peak searches.
CALCulate<Chn>:MARKer<Mk>:EXCursion:STATe
to activate it.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number
Parameters:
<Value>
Minimum peak excursion
The unit is derived from the active trace format and cannot be
modified. Changing the trace format resets the excursion to a
format-specific default value.
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:EXCursion:STATe
<Boolean>
Activates or deactivates the marker-specific minimum excursion for (single) peak
searches.
CALCulate<Chn>:MARKer<Mk>:EXCursion
to set the minimum peak excur-
sion.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number
Parameters:
<Active>
Boolean
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:FORMat
<OutFormat>
Sets/queries the output format for the (complex) value of the related marker.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number.