GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 889
Chapter 9
ERR-SUB-A-003.fm
122/047
DEBUG
ONLY
COMPRESSION
INVALID
LIFT_UP_PEDAL_E
RROR
Ignore this error. If the problem persists, contact your
OLC.
122/048
DEBUG
ONLY
COMPRESSION
INVALID
COMPRESSION_PE
DAL_ERROR
Ignore this error. If the problem persists, contact your
OLC.
122/049
DEBUG
ONLY
COMPRESSION
INVALID
DECOMPRESSION
_PEDAL_ERROR
Ignore this error. If the problem persists, contact your
OLC.
122/050
DEBUG
ONLY
COMPRESSION
INVALID
STOP_MOTION_DS
P_ACTIVE
Ignore this error. If the problem persists, contact your
OLC.
122/051
DEBUG
ONLY
COMPRESSION
INVALID
CLUTCH_MALFUN
CTION
Ignore this error. If the problem persists, contact your
OLC.
122/052
DEBUG
ONLY
COMPRESSION
INVALID
USERINTERFACE_
ERROR
Ignore this error. If the problem persists, contact your
OLC.
122/053
DEBUG
ONLY
COMPRESSION
INVALID
COLLIMATOR_BUT
TON_ERROR
Ignore this error. If the problem persists, contact your
OLC.
122/054
HALT
COMPRESSION
INVALID
RELAY_ERROR
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
122/055
DEBUG
ONLY
COMPRESSION
INVALID
UPPER_OPTICAL_
FORK_REACHED
Ignore this error. If the problem persists, contact your
OLC.
122/056
DEBUG
ONLY
COMPRESSION
INVALID
LOWER_OPTICAL_
FORK_REACHED
Ignore this error. If the problem persists, contact your
OLC.
122/057
DEBUG
ONLY
COMPRESSION
INVALID
INITIALZATION_ER
ROR
Ignore this error. If the problem persists, contact your
OLC.
122/058
HALT
COMPRESSION
INVALID
UNCONTROLLED_
MOTION
Ignore this error. If the problem persists, contact your
OLC.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card