GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A232 - Lift Screw Assembly
Page no. 1257
Chapter 9
JC-DR-A-232.fm
6
PROCEDURE
6-1
Disassembly of Gantry Curtains
1. Remove the two springs of curtains from the back of the Gantry. (Using a crook or tiewrap could
help). If necessary, move the trolley position so that the springs (1) that attach upper and lower cur-
tains are visible.
2. Cut the tiewrap fixing the clutch cable to curtain (2).
3. Release and remove the two lower screws securing the lower curtain (3) (3 mm allen wrench).
4. Remove the lower curtain.
1
2
3