Chapter 9
Page no. 866
ERR-SUB-A-003.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
111/001
DEBUG
ONLY
ROTATION SOFT-
WARE ERROR
On every command sent by the PTB, the IPMS
checks whether the command identifier is part of the
list of commands to be taken into account by the
IPMS. If the command is not recognized, it is
ignored, and “CommandIdError” is returned.
Should not occur in the field. It could be a disconnect
between the CPU soft release and the Node soft
release. If the problem persists, contact your OLC.
111/002
DEBUG
ONLY
ROTATION SOFT-
WARE ERROR
On every command sent by the PTB, the IPMS
checks whether all the parameter values are com-
patible with their corresponding range. If they are not
compatible, the IPMS ignores the command and
returns the “ParameterError” to the PTB.
Should not occur in the field. It could be a disconnect
between the CPU soft release and the Node soft
release. If the problem persists, contact your OLC.
111/003
WARNING
ROTATION CALIB
NOT ACHIEVED
If the calibration has not been achieved, for each
CAN message which is not a calibration command,
the IPMS ignores the command and returns the “Cal-
ibrationNotAchieved” error.
Perform
Job Card CAL A021 - Arm Rotation Calibra-
.
CAL A021
111/004
WARNING
NO VALID LINE
FOR ROTATION
Occurs when the Gantry CPU requests a motion but
node does not see the valid line. The valid line is an
RT line that is activated by the CPLD of the node
containing the UIF during the time when the UIF is
p a deceleration time. This activation is a
frequency put on the RT line and that is specific to
the button pressed (Tube Head and/or X-ray Con-
sole).
Also occurs when the valid line falls before the end of
the movement.
This error does not concern decompression and tilt.
Valid line problem. Check the different motions,
using all keypads to determine the defective node
then refer to
JC DIAG A1116 ROTATION CONTROL
for diagnostics.
DIAG
A1116
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card