GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card DIAG A001 - AWS Sun-VTS Diagnostics
Page no. 983
Chapter 9
JC-DIAG-A-001.fm
6. Once the test is complete reconnect the UPS cable to the serial port of the UPS.
7. Restart the UPS from the command line, by entering:
ups start
11-3-2 Full Tests without Cable on V3 Control Stations
1. Power up the system and wait for the
console login:
prompt on the AWS monitor.
2. Login as root as follows:
a. At the
console login:
prompt, login as root by entering:
root
b. When requested to specify the root password by entering:
operator
3. A list of start up messages appear, while the ADS Computer boots up. Wait for the following mes-
sage to appear:
*********************************************
* UPS Communication not established *
* System will shutdown *
*********************************************
4. Stop the UPS daemon, by entering:
/etc/rc2.d/S99PBEAgent stop
5. Once the UPS has stopped, remove the UPS cable from Slot #2 of PCI1 of the AWS Computer.
6. Plug a
Standard RS-232 Loopback Adapter
into Slot #2 of PCI1 on the AWS Computer.
7. Run the test from the command line, by entering:
tip /dev/term/3
The system answers
Connected
.
8. Type some alphanumeric characters on the keyboard. If the board is good, they are echoed immedi-
ately on the screen
9. To exit the test, type the command:
<Return>~
.
Note:
It can be necessary to repeat the sequence several times.
10. Once the test is complete reconnect the UPS cable to the serial port of the UPS.
11. Restart the UPS from the command line, by entering:
ups start
12
ADS-SENOGRAPHE LINE TESTS
12-1 Full Test without Cable
•
Type of test:
UNIX command
1. Power up the system and wait for the
console login:
prompt on the AWS monitor.
2. Login as root as follows:
a. At the
console login:
prompt, login as root by entering:
root
b. When requested to specify the root password by entering:
operator
3. Remove the Senographe cable from the port 1 of Digi Board (additional serial board.)