GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A015 - Lift Movement Troubleshooting
Page no. 1051
Chapter 9
JC-TSG-A-015.fm
ILLUSTRATION 8 - EXAMPLES OF GOOD AND BAD GREASING OF THE LIFT SCREW
3. Once you have finished applying grease to the entire Lift Screw, use the slow speed Arm Control but-
tons to drive the Lift over the whole range up and down two times in slow speed.
4. Use the fast speed Arm Control buttons to drive the Lift over the whole range up and down a few
times in fast speed.
5. Check how grease has been spread to the screw. Examples of well spread grease, smoothly and
uniformly along the whole screw length, with no splattered grease are shown in illustration 9.
Good
Uniformly applied grease.
Bad
Too much applied grease can splatter to
column wall. Clean surfaces, walls and
cables from spattered grease.