Chapter 9
Page no. 1020
JC-TSG-A-010.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A010 - Troubleshooting Bad Pixel Calibration Failure
ILLUSTRATION 1 - HIGH-LEVEL OVERVIEW OF THE BAD PIXEL CALIBRATION PROCESS
3-1
Troubleshooting - Failure at the End of IQ Check Image
Follow the appropriate sub-section below according to the IQ check test message you received.
3-1-1
IQ Check Test - Median Signal Level
The purpose of this test is to check that the system configuration is correct and that the correct phantom
has been placed on the detector. If this test fails, the calibration is aborted and the following message
appears:
"brightness level is not OK"
3-1-1-1 Typical Root Causes
1. The system is not correctly configured for bad pixel calibration (i.e. a compression paddle or a bucky
is in place).
2. The kV scale factor of the X-Ray tube is too high or too low.
3. The mAs measurement of the X-Ray tube is too high or too low.
4. The detector temperature is out of range.
3-1-1-2 Corrective Actions
1. Ensure that the bucky is removed, and that no compression paddle is installed. Then perform bad
pixel calibration with the flat field phantom as instructed.
2. If the problem persists, perform
in
Job Card CAL A011 - Calibration of
, to see if the kV Scale Factor is within the correct limits. If the kV Scale Factor results
fail, re-calibrate the kV scale factor according to
Job Card CAL A011 - Calibration of kV Scale Factor
. After kV scale factor calibration has been performed, go to
be performed after a kV scale factor calibration
to execute the other Job Cards in the
list.
3. If the problem persists, perform
Job Card CAL A009 - X-Ray Tube mA
, to see if the mAs of the X-Ray tube is within the correct limits. If the mAs results fail,
re-calibrate the mAs of the X - Ray tube according to
Job Card CAL A009 - X-Ray Tube mA Mea-
. After the mA calibration has been performed, go to
to be performed after an X-Ray tube mA measurement
to execute the other Job Cards
IQ check image:
First exposure
6 exposures
Display bad pixel
calibration results
Bad pixel calibration
can fail here
Bad pixel calibration
can fail here