GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A043 - AWS Monitor Calibration
Page no. 733
Chapter 7
JC-ELE-A-043.fm
5. If necessary, unlock the OSD (see
) and move the OSD to a convenient position (see
6. Re-enter the
Service level 2
menu again (as explained in steps
7. From the
Service level 2
menu, keep pressing the
MENU
button so select the
Calibration
menu, and
press
the
UP
button
to enter the
Calibration
menu.
8. In the
Calibration
menu, verify that the
Lut Backlight command
value is set to
OFF
and the
Select dis-
play function
is set to
4 LUT 1
.
If these values are set not as recommended:
•
change them by selecting it with the
MENU
button and change it to the correct value with the
UP
button
•
store the changed settings by pressing the
SET
button until
Accept changes
is selected, then
press the
UP
button to exit and save the settings.
9. Once the changed settings are saved, go to
for the 18" monitor.
INCORRECT
INCORRECT
CORRECT
CORRECT