Chapter 11
Page no. 1796
JC-PM-A-300.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card PM A300 - 21-inch 3MP Monitor Luminance Checks
7-2
Perform Luminance Tests
1. Ensure that the 21-inch 3MP Monitor has been powered on for at least 30 minutes. The 30 minute
warm up period is necessary before performing the luminance tests.
2. From the Patient List in the Browser, select the
AAPM Test Patterns
Patient.
3. In the Study list, select the
1K-LUMIN
Study ID.
4. In the Series list, select Series
2
(Procedure Description
TG18-LN
).
5. Click the
Review Exam
icon. The
Exam Review
window appears with images from the series.
6. If not already selected, use the
Img
up/down
buttons to select the first image (number
1
) in the
series. Ensure that the title of the displayed image is
TG18-LN12-01 Pattern
.
AAPM Test Patterns
Lumiance Meter
Position