Chapter 9
Page no. 1726
JC-CAL-A-053.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card CAL A053 - mAs Non-linearity Calibration
9. Exit the Calibration Menu (
SETUP/SETUP/SETUP//SETUP/SETUP
).
6-3
Acceptable Values
•
The value displayed for MASPRE CORR should be within the limits 1.00 and 1.20.
•
If it is outside this range, check the Generator calibration and repeat the calibration.
7
COMPLETION
If you are handing the machine back to the user at this point, switch the Senographe off, then on again.