Chapter 9
Page no. 1328
JC-DR-A-250.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A250 - Compression Board
6-4
Observation LED Checks
•
Power on the Gantry and check that no errors are raised on the Gantry LCD display.
•
Check that the LED states are as follows. For more information about the behavior and states of the
Compression board LEDs, refer to
.
LED Label
LED Color
LED Status
DS1
Green
ON
DS2
Green
ON
DS3
Yellow
ON
DS4
Yelllow
ON
DS5
Green
ON
DS6
Yellow
OFF
DS7
Green
ON or OFF
DS8
Yellow
ON
DS9
Green
ON
DS10
Red
Blinking