Chapter 9
Page no. 982
JC-DIAG-A-001.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card DIAG A001 - AWS Sun-VTS Diagnostics
11-2-2 Full Tests without Cable on V3 Control Stations
1. Power up the system and wait for the
console login:
prompt on the AWS monitor.
2. Login as root as follows:
a. At the
console login:
prompt, login as root by entering:
root
b. When requested to specify the root password by entering:
operator
3. A list of start up messages appear, while the ADS Computer boots up. Wait for the following mes-
sage to appear:
*********************************************
* UPS Communication not established *
* System will shutdown *
*********************************************
4. Stop the UPS daemon, by entering:
/etc/rc2.d/S99PBEAgent stop
5. Once the UPS has stopped, remove the UPS cable from Slot #2 of PCI1 of the AWS Computer.
6. Plug a
Standard RS-232 Loopback Adapter
into Slot #2 of PCI1 on the AWS Computer.
7. Run the test from the command line, by entering:
tip /dev/term/3
The system answers
Connected
.
8. Type some alphanumeric characters on the keyboard. If the board is good, they are echoed immedi-
ately on the screen
9. To exit the test, type the command:
<Return>~
.
Note:
It can be necessary to repeat the sequence several times.
10. Once the test is complete reconnect the UPS cable to Slot #2 of PCI1of the AWS Computer.
11. Restart the UPS from the command line, by entering:
ups start
11-3 Full Test with Cable
•
Type of test:
SUN VTS
11-3-1 Full Tests with Cable on V1/V2 Control Stations
1. Power up the system and wait for the
console login:
prompt on the AWS monitor.
2. Within 20 seconds of seeing the login prompt, stop the UPS daemon by doing the following:
a. At the
console login:
prompt, login as root by entering:
root
b. When requested to specify the root password, enter:
operator
c. Stop the UPS daemon, by entering:
ups stop
3. Once the UPS has stopped, remove the UPS cable from the serial port of the UPS.
4. Plug a loopback connector "
" into the end of the disconnected UPS cable.
5. Run the test from the command line, by entering:
/opt/SUNWvts/bin/sptest –v -f -o dev=se0,L=P,T=b