Chapter 9
Page no. 1582
JC-DR-A-407.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A407 - Generator Interface Board 400-PL2
7
COMPLETION
Carry out the following calibrations after you have changed the Generator Interface board:
Calibrations - Checks
Job Card
Bias Calibration
Job Card CAL A003 - X-Ray Tube Focal Bias Voltage Scale Factor
Heating Current Calibration
Job Card CAL A001 - Heater Current Scale Factor
kV Scale Factor
Job Card CAL A011 - Calibration of kV Scale Factor
mAs X-ray Calibration
Job Card CAL A012 - X-Ray Tube Heater Current
AOP/SNR Check