GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 873
Chapter 9
ERR-SUB-A-003.fm
112/035
HALT
ROTATION TX CAN
BUFFER OVERRUN
The transmit queue between the MCU and the DSP
is full and message dropped.
Change the Rotation board then perform a Rotation
Calibration.
D/R A221
CAL 021
112/036
HALT
ROTATION MAIL-
BOX OVERRUN
The mailbox queue used in the inter-task communi-
cation in the MCU is full and mails dropped.
Change the Rotation board then perform a Rotation
Calibration.
D/R A221
CAL 021
112/037
HALT
CPU HEARTBEAT
NOT RECEIVED BY
ROTATION NODE
There has been a heartbeat failure. The Gantry CPU
did not send periodic "heartbeat" CAN message.
1. Change the Gantry CPU Board (see
2. Perform a Collimator Calibration (see
CAL A029 - Collimator Format Calibration
).
3. Perform a Collimation Check (see
D/R A190
CAL A029
ELE A028
112/038
DEBUG
ONLY
HEARTBEAT DIS-
ABLED ON ROTA-
TION NODE
The Gantry CPU tried to disable the "heartbeat"
check of the rotation node.
1. Change the Gantry CPU Board (see
2. Perform a Collimator Calibration (see
CAL A029 - Collimator Format Calibration
).
3. Perform a Collimation Check (see
D/R A190
CAL A029
ELE A028
112/039
HALT
ROTATION INVALID
PDO LENGTH
The rotation node receives a PDO with an invalid
DLC.
Refer to
JC DIAG A1117 ROTATION COMMUNICA-
to fix the problem.
A1117
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card